For the magnetic susceptibility measurement, two types of magnetic susceptibility
meters were used, i.e. gModel 3101Ah(Bison Instruments, U.S.A) and gModel
MS 2h(Bartington Instruments, U.K).
For the natural remanent magnetization (NRM) measurement, a spinner magnetometer hModel SMD- 88 (SMM-85)h (Natsuhara Gikken, Japan) was usedD
To determine the Curie temperature, a type of gVSM-155g(Princeton Applied Research Corp., U.S.A) was used for the measurement of thermomagnetic curve in the magnetic field of 0.8 T under 10-3 Pa in vacuum.
Dry density measurement was conducted after drying the specimen in the
air at room temperature for several months. Wet density measurement using
deionized water was made for the specimen left in the vacuum chamber with
water after 40 hours.
Ultrasonic velocity (Vp) was measured using the combination of gpulse generator and syncroscopeh with a 0.5 or 1 MHz transducer. Or an ultrasonic velocity measurement device, gModel NSPG-53Dh(Kaihatsu Densisokki, Japan) under the room temperature and atmospheric pressure.